- input type thermocouple: R, S, PR40-20, Au-Fe, K, E, J, T, C, N, PLII, U, L, W5Re, W26Re
- input type RTD: PT100 or JPT100
- input type DC voltage: mV or V
- input type DC current: 4-20mA (with extra external shunt resistor)
- number of internal inputs: 3, 6 or 12 (isolated)
- number of external inputs: max. 6 pieces of 6 channel external analog input (via MODBUS RTU RS-485 interface). Resulting in max. 48 channel recording system.
- fastest measurement cycle: 100msec.
- data recording cycle: 1 sec to 60 min (100ms is selectable only for sub-recording)
- display unit: 5.7″ color TFT with touchscreen operation
- external storage: SD memory card (max. 32Gb) SDHC
- file storage cycle: 1 hour to 1 year
- alarm function: up to 4 points per channel
- alarm type: high limit, low limit, abnormal data
- alarm output: overall output, 1 point (open collector)
- MODBUS TCP ethernet interface (with built-in HTTP & FTP server)
- power supply: 100-240Vac
- dustproof & waterproof rating: IP65 (front panel only)
- optional
- extended security for FDA 21 CFR part 11
- MODBUS RTU RS-485 interface
- digital inputs (9 points dry contact input i.e. RUN/STOP, print message, backlight on/off)
- digital outputs (12 points open collector output or 6 points relay contact output for alarm)
- built inside a rugged stainless steel housing in RAL color to your specification
- calibration certificate
– Leaflet VGR-B100
– Operation manual VGR-B100
– Data viewer manual VGR-B100
– Extended security manual VGR-B100 for FDA 21 CFR part 11
– Parameter loader manual VGR-B100
– Communication manual VGR-B100
– Remote analog input manual VGR-B100
– Download data viewer & parameter loader VGR-B100 (click on “Software” tab)
Click here to request a quotation for the VGR-B100.




Recording starts immediately after pressing the REC button

Data is stored in the SD card. With the supplied software, the recorded data can be easily viewed and managed.
Various input modules

The number of channels can be easily extended to max. 48 channels with optional 6 channel analog input modules (communication via MODBUS RTU RS-485 interface).
Extended security function for conformity with FDA 21 CFR part 11 (optional)

- Access limit: only users with a registered account can operate the recorder
- Audit trail: operation history such as change of set values, recording start/stop and so on is automatically recorded
- Electronic signature, signature data output: both trend and instantaneous values with electronic signature data can be printed out
- Data falsification and damage are detected: viewing detect data will be limited
External dimensions

Stainless steel housing

Can be built inside a rugged stainless steel housing in RAL color to your specification
Model codes